In 488 Outpatients, Anterior Nasal and Joint Nose-Throat Sampling Agreed 98% to 99% for Three Respiratory Viruses.

Joint nose and throat sampling can improve anatomical coverage, but it also adds discomfort, training, and handling. A Hong Kong study collected paired joint nose-throat swabs and a novel anterior nasal specimen from 488 outpatients with febrile acute respiratory illness. RT-qPCR tested influenza A, influenza B, SARS-CoV-2, and human RNase P as a collection control.

A healthcare worker collecting a nasal swab specimen
Nasal swab collection. Claude Truong-Ngoc, CC BY-SA 4.0. Image source.

Agreement was high under supervised collection

Overall agreement was 98.0% for influenza A and 99.0% for both influenza B and SARS-CoV-2. Cohen's kappa was 0.95, 0.94, and 0.92, respectively. Viral and RNase P Ct values showed moderate correlation between specimens. These results support anterior nasal collection as a viable alternative in a supervised outpatient setting. Stability should also be challenged at the warmest transport condition and longest claimed holding time.

Agreement can be inflated when most results are negative, so it should not replace positive agreement or clinical sensitivity. Sixteen of 18 discordant pairs had Ct values of at least 30. Low-load specimens therefore define the edge: small changes in rotation, dwell time, illness day, or transport may move a result across the cutoff. Lot acceptance can include weak human-control and low-positive materials.

Convenience should be measured, not assumed

An anterior nasal device normally avoids deep nasopharyngeal insertion and can fit repeated testing. A procurement study should still record pain, tearing, sneezing, refusal, collection time, and operator retraining. “Minimally invasive” has operational value only when it reduces failed collections or improves participation.

Instructions must define insertion depth, direction, rotations per nostril, and dwell time. RNase P or another human control can flag insufficient collection. Children, older adults, people with dry nasal mucosa, and patients early in illness may differ from the combined estimate. Performance from supervised collection cannot simply be assigned to unsupervised self-collection.

Low positives should anchor equivalence testing

A validation panel should deliberately contain specimens near the detection limit for all three viruses. Discordant pairs can be repeated or resolved with a third method or additional anatomical site. Invalid results belong in the denominator. Excess transport-medium volume should also be avoided because it dilutes limited anterior nasal material.

Changfeng Medical can supply anterior nasal, nasopharyngeal, and oropharyngeal flocked swabs, dual-site collection kits, and molecular-compatible transport tubes. Target-population testing can compare internal-control Ct, low-positive recovery, comfort, and hands-on time. Agreement of 98% to 99% is encouraging; the final substitution claim must come from the intended workflow.

Source: diagnostic agreement study of anterior nasal and joint nose-throat specimens.